Type de publicationJournal Article
Year of Publication2018
AuteursWu Y, Le Gorrec Y
JournalIFAC PAPERSONLINE
Volume51
Pagination199-204
Type of ArticleProceedings Paper
ISSN2405-8963
Mots-cléslinear quadratic optimization, Medical endoscope, optimal actuator location, Port-Hamiltonian system
Résumé

This paper deals with optimal actuator location for a medical endoscope controlled by electro-active polymer (EAP). The inner tube of the endoscope is a flexible structure that can be represented by a Timoshenko beam. Actuators are patches of EAP. There is freedom in the choice of EAP actuators location. In this paper, we first propose a port Hamiltonian model of the endoscope. In order to choose the optimal location for the EAP actuators, we consider the linear quadratic (LQ) performance as the optimal performance objective. At last, some numerical simulation results are given based on the real experimental setup parameters. (C) 2018, IFAC (International Federation of Automatic Control) Hosting by Elsevier Ltd. All rights reserved.

DOI10.1016/j.ifacol.2018.06.053