Label-free detection of pepsinogen 1 and 2 by polyethylene coating Lamb microfluidic device

Affiliation auteurs!!!! Error affiliation !!!!
TitreLabel-free detection of pepsinogen 1 and 2 by polyethylene coating Lamb microfluidic device
Type de publicationJournal Article
Year of Publication2019
AuteursWei W, Zhang W, Li C, Kong H, Guo Z, Zhang Z, Bastien F, Gong Y, Wang H, Zhou L
JournalBIOSENSORS & BIOELECTRONICS
Volume129
Pagination231-237
Date PublishedMAR 15
Type of ArticleArticle
ISSN0956-5663
Mots-clésGastric cancer, Immunoassay, Lamb sensor, Pepsinogens
Résumé

Early screening of gastric cancer is a critical importance for the improvement of patients' survival rate. Here, a polyethylene coating Lamb (PE-Lamb) microfluidic device with immune layer for gastric cancer label-free detection was constructed. Two serum pepsinogen 1 (PG1) and pepsinogen 2 (PG2) biomarkers were applied to screen and predict the appearance of gastric cancer. Compared with enzyme-linked immunosorbent assay (ELISA), this method achieved a higher sensitivity and less time (40 min vs 120 min). The limit of detections (LOD) were reached 60 pg/mL for PG1 and 30 pg/mL for PG2, which have two orders of magnitude lower than traditional ELISA. The linearity coefficient indexes (R-2) for PG1 and PG2 were 0.992 and 0.953 respectively, which is similar to that of ELISA. In addition, PG1 and PG2 mixed antigens sample with human serum was detected by PE-Lamb approach, and the frequency response showed high reproducibility and specificity. The results indicate that PE-lamb diagnostic technique is a novel and promising method for high-throughput screening and early diagnosis of gastric cancer.

DOI10.1016/j.bios.2018.09.075