Performance of two new automated assays for measuring von Willebrand activity: HemosIL Acu Star and Innovance

Affiliation auteurs!!!! Error affiliation !!!!
TitrePerformance of two new automated assays for measuring von Willebrand activity: HemosIL Acu Star and Innovance
Type de publicationJournal Article
Year of Publication2014
AuteursDe Maistre E, Volot F, Mourey G, Aho LS, Ternisien C, Briquel M-E, Bertrand M-A, Tardy B, Frotscher B, Nguyen P, Dumont L, Vandroux D, Hezard N, Trossaert M
JournalTHROMBOSIS AND HAEMOSTASIS
Volume112
Pagination825-830
Date PublishedOCT
Type of ArticleArticle
ISSN0340-6245
Mots-clésaggregometry, ristocetin cofactor activity, von Willebrand disease, Willebrand activity assay
Résumé

The ristocetin cofactor activity assay (VWF:RCo) is the reference method for assessing von Willebrand factor (VWF) activity but remains difficult to perform, and the coefficient of variation of the method is high (about 20-30%). This study evaluated and compared the performance for measuring the VWF activity of two newly commercialised assays [VWF:Ac Innovance (VWF:Ac) and VWF:RCo Acustar (VWF:RCo Acu)] with the reference VWF:RCo aggregation in 123 pathological plasma samples. The correlation and concordance between both new tests (VWF:RCo-Acu and VWF:Ac) and the reference VWF:RCo were good. The results of the VWF activity to VWF antigen ratio were also comparable whatever the method for the classification of VWF deficiency in all patients. Our results showed that both new tests could replace the ``gold standard'' VWF:RCo in aggregometry with several benefits: they are fully automated, easier and faster to perform, better adapted to emergency situations if necessary.

DOI10.1160/TH14-02-0108