Framework for patient-ventilator asynchrony during long-term non-invasive ventilation
Affiliation auteurs | !!!! Error affiliation !!!! |
Titre | Framework for patient-ventilator asynchrony during long-term non-invasive ventilation |
Type de publication | Journal Article |
Year of Publication | 2019 |
Auteurs | Gonzalez-Bermejo J, Janssens J-P, Rabec C, Perrin C, Lofaso F, Langevin B, Carlucci A, Lujan M, Grp SNIV |
Journal | THORAX |
Volume | 74 |
Pagination | 715-717 |
Date Published | JUL |
Type of Article | Article |
ISSN | 0040-6376 |
Résumé | Episodes of patient-ventilator asynchrony (PVA) occur during acute and chronic non-invasive positive pressure ventilation (NIV). In long-term NIV, description and quantification of PVA is not standardised, thus limiting assessment of its clinical impact. The present report provides a framework for a systematic analysis of polygraphic recordings of patients under NIV for the detection and classification of PVA validated by bench testing. The algorithm described uses two different time windows: rate asynchrony and intracycle asynchrony. This approach should facilitate further studies on prevalence and clinical impact of PVA in long-term NIV. |
DOI | 10.1136/thoraxjnl-2018-213022 |