Removal of evidential motion-contaminated and poorly fitted image data improves IVIM diffusion MRI parameter scan-rescan reproducibility

Affiliation auteurs!!!! Error affiliation !!!!
TitreRemoval of evidential motion-contaminated and poorly fitted image data improves IVIM diffusion MRI parameter scan-rescan reproducibility
Type de publicationJournal Article
Year of Publication2018
AuteursChevallier O, Zhou N, He J, Loffroy R, Wang YXiang J
JournalACTA RADIOLOGICA
Volume59
Pagination1157-1167
Date PublishedOCT
Type of ArticleArticle
ISSN0284-1851
Mots-clésDiffusion-weighted imaging, intravoxel incoherent motion, Liver, reproducibility
Résumé

{Background: It has been reported that intravoxel incoherent motion (IVIM) diffusion magnetic resonance imaging (MRI) scan-rescan reproducibility is unsatisfactory. Purpose: To study IVIM MRI parameter reproducibility for liver parenchyma after the removal of motion-contaminated and/or poorly fitted image data. Material and Methods: Eighteen healthy volunteers had liver scans twice in the same session to assess scan-rescan repeatability, and again in another session after an average interval of 13 days to assess reproducibility. Diffusion-weighted images were acquired with a 3-T scanner using respiratory-triggered echo-planar sequence and 16 b-values (0-800 s/mm(2)). Measurement was performed on the right liver with segment-unconstrained least square fitting. Image series with evidential anatomical mismatch, apparent artifacts, and poorly fitted signal intensity vs. b-value curve were excluded. A minimum of three slices was deemed necessary for IVIM parameter estimation. Results: With a total 54 examinations, six did not satisfy inclusion criteria, leading to a success rate of 89%, and 14 volunteers were finally included for the repeatability/reproducibility study. A total of 3-10 slices per examination ( mean = 5.3 slices

DOI10.1177/0284185118756949