Combined effects of leaks, respiratory system properties and upper airway patency on the performance of home ventilators: a bench study

Affiliation auteurs!!!! Error affiliation !!!!
TitreCombined effects of leaks, respiratory system properties and upper airway patency on the performance of home ventilators: a bench study
Type de publicationJournal Article
Year of Publication2017
AuteursZhu K, Rabec C, Gonzalez-Bermejo J, Hardy S, Aouf S, Escourrou P, Roisman G
JournalBMC PULMONARY MEDICINE
Volume17
Pagination145
Date PublishedNOV 21
Type of ArticleArticle
ISSN1471-2466
Mots-clésNon-invasive ventilation, Patient-ventilator asynchrony, Unintentional leak, Upper airway patency
Résumé

Background: Combined effects of leaks, mechanical property of respiratory system and upper airway (UA) patency on patient-ventilator synchrony (PVA) and the level of clinically ``tolerable'' leaks are not well established in home ventilators. Methods: We comparatively assessed on a bench model, the highest leak level tolerated without inducing significant asynchrony (''critical leak'') in three home ventilators (Astral 150, Trilogy 100 and Vivo 60; noted as A150, T100 and V60 respectively) subjected to three simulated diseased respiratory conditions: chronic obstructive pulmonary disease (COPD), obesity hypoventilation (OHS) and neuromuscular disorders (NMD), with both open and closed UA. Also, total leak values in the device reports were compared to the bench-measured values. Results: With open UA, all ventilators were able to avoid asynchrony up to a 30 L/min leak and even to 55 L/min in some cases. UA closure and respiratory diseases especially OHS influenced PVA. With closed UA, the critical leak of A150 and T100 remained higher than 55 L/min in COPD and OHS, while for V60 decreased to 41 and 33 L/min respectively. In NMD with closed UA, only T100 reached a high critical leak of 69 L/min. Besides, inspiratory trigger sensitivity change was often necessary to avoid PVA. Conclusions: Home ventilators were able to avoid PVA in high-level leak conditions. However, asynchrony appeared in cases of abnormal mechanical properties of respiratory system or closed UA. In case of closed UA, the EPAP should be adjusted prior to the inspiratory trigger.

DOI10.1186/s12890-017-0487-2