Transaortic transcatheter aortic valve implantation using SAPIEN XT or SAPIEN 3 valves in the ROUTE registry
Affiliation auteurs | !!!! Error affiliation !!!! |
Titre | Transaortic transcatheter aortic valve implantation using SAPIEN XT or SAPIEN 3 valves in the ROUTE registry |
Type de publication | Journal Article |
Year of Publication | 2017 |
Auteurs | Romano M, Frank D, Cocchieri R, Jagielak D, Bonaros N, Aiello M, Lapeze J, Laine M, Chocron S, Muir D, Eichinger W, Thielmann M, Labrousse L, Rein KArne, Verhoye J-P, Gerosa G, Baumbach H, Deutsch C, Bramlage P, Thoenes M, Bapat V |
Journal | INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY |
Volume | 25 |
Pagination | 757-764 |
Date Published | NOV |
Type of Article | Article |
ISSN | 1569-9293 |
Mots-clés | Balloon expandable, Mortality, transaortic, transcatheter aortic valve implantation |
Résumé | OBJECTIVES: Transaortic (TAo) access for transcatheter aortic valve implantation (TAVI) is an alternative to the conventional transfemoral or transapical routes. Data comparing the characteristics and outcomes of TAo-TAVI using the SAPIEN XT and SAPIEN 3 heart valves are scarce. The objective of the current analysis was to provide such information. METHODS: ROUTE is an international, prospective, observational registry. Patients with severe calcific aortic stenosis scheduled for TAo-TAVI with an Edwards SAPIEN XT or a SAPIEN 3 heart valve were consecutively enrolled at 22 centres across Europe between February 2013 and February 2015. Periprocedural, in-hospital and 30-day complication rates were assessed. RESULTS: Of the 301 patients included, 126 (41.9%) received a SAPIEN 3 and 175 (58.1%) a SAPIEN XT. The SAPIEN 3 was associated with shorter procedure time (101 +/- 35 vs 111 +/- 40 min; P = 0.031) and a lower quantity of contrast agent used (87 +/- 43 vs 112 +/- 50 ml; P < 0.001). Balloon dilation was performed less often before (68.0% vs 78.3%; P = 0.045) and after implantation (13.6% vs 30.1%; P = 0.001). No statistically significant differences between the valve types were documented for overall (4.1% SAPIEN 3 vs 7.6% SAPIEN XT; P = 0.21), TAVI-related (0.8% vs 4.7%; P = 0.084) and cardiovascular mortality (2.4% vs 5.9%; P = 0.158). Major vascular complications were less frequent (0.8% vs 5.3%; P = 0.049), and there was a lower rate of moderate-to-severe paravalvular regurgitation (0.8% vs 5.1%; P = 0.050) in the SAPIEN 3 group. CONCLUSIONS: Both the SAPIEN XT and SAPIEN 3 were safely implanted via the TAo route, though the SAPIEN 3 may be associated with a higher procedural success rate and improved prognosis. |
DOI | 10.1093/icvts/ivx159 |