Accuracy in obtaining 100 mu g from 10 mg of morphine for spinal anesthesia
Affiliation auteurs | !!!! Error affiliation !!!! |
Titre | Accuracy in obtaining 100 mu g from 10 mg of morphine for spinal anesthesia |
Type de publication | Journal Article |
Year of Publication | 2015 |
Auteurs | Benkhadra M, Rivory J-C, Wessels C, Guerard P, Vadot L, Astruc K, Dumas M, Girard C |
Journal | JOURNAL OF CLINICAL ANESTHESIA |
Volume | 27 |
Pagination | 638-645 |
Date Published | DEC |
Type of Article | Article |
ISSN | 0952-8180 |
Mots-clés | Analgesia, anesthesia, Morphine |
Résumé | Study objective: Dilution is often required to obtain appropriate concentrations of intrathecal morphine for analgesia. We compared techniques of diluting by measuring the quantity of morphine actually obtained in the final solution. Design: This is an experimental study by 3 experienced anesthesiologists. Setting: The setting is at a university teaching hospital. Patients: There are no patients. Interventions: There are no interventions. Measurements: Five techniques for obtaining 100 jig from 10 mg/mL were compared: technique 1 (T1) = extraction up to 0.1 graduation on a 1-mL syringe, followed by simple dilution (SD). Technique 2 (T2) = As for T1 but syringe was shaken to mix solution. Technique 3 (T3): SD with 10-mL syringe. Technique 4 (T4): Double dilution with 10-mL syringe. Technique 5 (T5): Extraction up to the 0.1 graduation of a 1-mL syringe, then SD, then shake solution by hand. Three tests using high-performance liquid chromatography with ultraviolet were performed on each syringe prepared 3 consecutive times, namely, at the first (beginning, B), fifth (middle, M) and last (end, E) milliliter or 0.1 mL (depending on syringe type). Main results: Average overall concentrations were 208 19, 199 24, 120 13, 136 9, and 119 16 mu g/0.1 mL, T1-T5, respectively. By Kruskal-Wallis test, we classified the techniques according to the magnitude of the difference between the observed concentration of morphine and the desired (theoretical) concentration of 100 mu g/0.1 mL. In ascending order, techniques ranked as follows: TS (smallest difference), T3, T4, T2, and T1 (greatest difference) (P = .0001). Conclusions: There is significant variability in the concentration of morphine actually contained in final solutions after dilution. Morphine presented in different premixed concentrations increases the risk of error. We advocate technique 5 as described above, whereas technique 1 should be prohibited. (C) 2015 Elsevier Inc. All rights reserved. |
DOI | 10.1016/j.jclinane.2015.06.004 |