Conception and realization of a parallel-plate free-air ionization chamber for the absolute dosimetry of an ultrasoft X-ray beam

Affiliation auteurs!!!! Error affiliation !!!!
TitreConception and realization of a parallel-plate free-air ionization chamber for the absolute dosimetry of an ultrasoft X-ray beam
Type de publicationJournal Article
Year of Publication2014
AuteursGroetz J.-E, Ounoughi N., Mavon C., Belafrites A., Fromm M.
JournalREVIEW OF SCIENTIFIC INSTRUMENTS
Volume85
Pagination083304
Date PublishedAUG
Type of ArticleArticle
ISSN0034-6748
Résumé

We report the design of a millimeter-sized parallel plate free-air ionization chamber (IC) aimed at determining the absolute air kerma rate of an ultra-soft X-ray beam (E = 1.5 keV). The size of the IC was determined so that the measurement volume satisfies the condition of charged-particle equilibrium. The correction factors necessary to properly measure the absolute kerma using the IC have been established. Particular attention was given to the determination of the effective mean energy for the 1.5 keV photons using the PENELOPE code. Other correction factors were determined by means of computer simulation (COMSOL (TM) and FLUKA). Measurements of air kerma rates under specific operating parameters of the lab-bench X-ray source have been performed at various distances from that source and compared to Monte Carlo calculations. We show that the developed ionization chamber makes it possible to determine accurate photon fluence rates in routine work and will constitute substantial time-savings for future radiobiological experiments based on the use of ultra-soft X-rays. (C) 2014 AIP Publishing LLC.

DOI10.1063/1.4890817