Determination of plutonium in nitric acid solutions using energy dispersive L X-ray fluorescence with a low power X-ray generator
Affiliation auteurs | !!!! Error affiliation !!!! |
Titre | Determination of plutonium in nitric acid solutions using energy dispersive L X-ray fluorescence with a low power X-ray generator |
Type de publication | Journal Article |
Year of Publication | 2015 |
Auteurs | Py J., Groetz J.-E, Hubinois J.-C, Cardona D. |
Journal | NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT |
Volume | 780 |
Pagination | 131-137 |
Date Published | APR 21 |
Type of Article | Article |
ISSN | 0168-9002 |
Mots-clés | Compton Peak correction, EDXRF, L X-rays, Plutonium, SDD detector, Secondary target |
Résumé | This work presents the development of an in line energy dispersive LX-ray fluorescence spectrometer setup, with a low power X-ray generator and a secondary target, for the determination of plutonium concentration in nitric acid solutions. The intensity of the L X-rays from the internal conversion and gamma rays emitted by the daughter nuclei from plutonium is minimized and corrected, in order to eliminate the interferences with the L X-ray fluorescence spectrum. The matrix effects are then corrected by the Compton peak method. A calibration plot for plutonium solutions within the range 0.1-20 g L-1 is given. (C) 2015 Elsevier B.V. All rights reserved. |