Sensitivity enhancement of a fluorescent pH sensor by double silanization of the sensing surface
Affiliation auteurs | !!!! Error affiliation !!!! |
Titre | Sensitivity enhancement of a fluorescent pH sensor by double silanization of the sensing surface |
Type de publication | Journal Article |
Year of Publication | 2019 |
Auteurs | Gauthier-Manuel B, Kateklum R, Pieralli C, Mankhetkorn S, Wacogne B |
Journal | MATERIALS TODAY-PROCEEDINGS |
Volume | 6 |
Pagination | 345-349 |
Type of Article | Proceedings Paper |
ISSN | 2214-7853 |
Mots-clés | APDMS, APTMS, Fluorescence, pH sensor, Silanization |
Résumé | The goal of this work concerns a new method to substantially increase the sensitivity of a fluorescence pH sensor. The method is based on a double silanization method. Two methods are compared: one using APTMS (3-Aminopropyl)trimethoxysilane) only and another one using both APTMS and APDMS (3-Aminopropyl)dimethoxymethylsilane). Using the second method, sensor's sensitivity is improved by more than 500 %. (C) 2018 Elsevier Ltd. All rights reserved. |
DOI | 10.1016/j.matpr.2018.10.429 |