Well GeHP detector calibration for environmental measurements using reference materials

Affiliation auteurs!!!! Error affiliation !!!!
TitreWell GeHP detector calibration for environmental measurements using reference materials
Type de publicationJournal Article
Year of Publication2016
AuteursTedjani A., Mavon C., Belafrites A., Degrelle D., Boumala D., Rius D., Groetz J.-E
JournalNUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
Volume838
Pagination12-17
Date PublishedDEC 1
Type of ArticleArticle
ISSN0168-9002
Mots-clésEfficiency correction, Full energy peak efficiency, MCNP6, True coincidence summing, Well-type detector
Résumé

A well-type detector installed in the Modane underground Laboratory (LSM) can combine both low background and high detection efficiency and it is well suited for the analysis of small amounts of environmental samples. Reference materials such as IAEA-447 (moss-soil), IAEA-RG-Thl and IAEA-RG-U1 were used for the detector calibration, owing to a chemical composition close to those of the environmental samples. Nevertheless, the matrix effects and the true coincidence summing effects must be corrected from the full energy peak efficiency (FEPE). The FEPE was performed for a wide range of energy by a semi-empirical method using Monte Carlo simulation (MCNP6), intended for environmental measurements such as lake sediments dating. In the well geometry, the true coincidence summing effects could be very important and correction factors have been computed in three different ways. (C) 2016 Elsevier B.V. All rights reserved.

DOI10.1016/j.nima.2016.09.022